Oxygen Barrier Tester ASTM D3985 is applicable to the determination of oxygen transmission rate of:
Basic Applications | Films Including plastic films, plastic composite films, paper-plastic composite films, geomembranes coextruded films, aluminized films, aluminum foil, aluminum composite films, and many others |
Sheeting Including various sorts of engineering plastics, rubber, and building materials, e.g. PP, PVC and PVDC |
|
Packages Including plastics, rubber, paper, paper-plastic composite, glass, and metal packages, e.g. Coke bottles, peanut oil packages, Tetra Park materials, vacuum bags, metal three-piece cans, plastic packages for cosmetic, soft tube for toothpaste, jelly and yogurt cups |
Specifications:
Oxygen Barrier Tester ASTM D3985
1. Test 3 equivalent specimens
2. 2 test modes
3. Conform to ASTM D3985
4. CE,ISO9001
Oxygen Barrier Tester ASTM D3985 is based on the equal pressure method, and is professional applicable to the determination of oxygen transmission rate of film and package products, including plastic films, composite films, sheeting, plastic bottles, plastic bags and other packages.
1. Test 3 equivalent specimens at one operation and export test results in average value
2. 2 test modes for both films and packages, and accessories for package test are available for customization
3. Equipped with constant temperature and humidity control devices (optional parts) to meet different test requirements
4. Reference film for fast calibration to ensure accurate and universal test data
5. The instrument is controlled by micro-computer with LCD, menu interface and PVC operating panel, which could conveniently export test data, test results and test curves
Standards:
ISO 15105-2, GB/T 19789, ASTM D3985, ASTM F2622, ASTM F1307, ASTM F1927, JIS K7126-2, YBB00082003
Technical Parameter:
Specifications | Film Test | Package Test (customization available) |
Test Range | 0.01~1000 cm3/m2·d (standard) 0.1~10,000 cm3/m2·d (optional) |
0.0001~10 cm3/pkg·d (standard) |
Number of Specimens | 1~3 | |
Resolution | 0.01 cm3/m2·d | 0.0001 cm3/pkg·d |
Temperature Range | 15°C~55°C (not in supply scope) | / |
Temperature Accuracy | ±0.1°C | / |
Humidity Range | 0%RH, 15%RH~90%RH, 100%RH (not in supply scope) |
|
Humidity Accuracy | ±1%RH | |
Test Gas | O2 and air (not in supply scope) | |
Test Area | 50 cm2 | / |
Thickness | ≤3 mm (customization available) | / |
Specimen Size | 108 mm x 108 mm | Test in 100% O2: Specimen should be smaller than 120mm and lower than 360mm |
Test in the Air: No size limitation | ||
Bottles: the inner diameter should be bigger than 8mm, outer diameter should be smaller than 42mm (Standard) | ||
Bags or Boxes: Supported by accessories | ||
Carrier Gas | High purity nitrogen with more than 99.999% concentration (not in supply scope) |
|
Gas Supply Pressure | ≥ 0.28MPa | |
Port Size | 1/8 inch copper tubing | |
Instrument Size | 670mm(L) x 410mm(W) x 310mm(H) | 670mm(L) x 410mm(W) x 630mm(H) |
Power | AC (85~264)V (47~63)Hz | |
Net Weight | 48kg | 50kg |