Send Message
Home > products > Flammability Testing Equipment >
ISO 1133 ASTM D1238 Melt Flow Index Tester , MFR/MVR Mfi Testing Machine

ISO 1133 ASTM D1238 Melt Flow Index Tester , MFR/MVR Mfi Testing Machine

ASTM D1238 melt flow index tester

mfi melt flow index tester

ISO 1133 mfi testing machine

Place of Origin:

China

Brand Name:

YUYANG

Model Number:

YYW098

Contact Us
Request A Quote
Product Details
Product Name:
Melt Flow Index Test
Function:
MFR/MVR
Temperature Range:
Room Temperature - 400°C
Temperature Accuracy:
±0.2°C
Temperature Resolution:
0.1°C
Timing Range:
0.1-999 S
Timing Accuracy:
0.1s
Instrument Dimensions:
450×400×610(L × W × H)
Highlight:

ASTM D1238 melt flow index tester

,

mfi melt flow index tester

,

ISO 1133 mfi testing machine

Payment & Shipping Terms
Minimum Order Quantity
1PCS
Price
$1150
Packaging Details
Wooden Box
Delivery Time
7-30 Days
Payment Terms
T/T, Western Union, MoneyGram
Product Description

ISO 1133 ASTM D1238 Melt Flow Index Tester

 

Description:

The melt mass-flow rate (MFR) and the melt volume-flow rate (MVR) are determined by extruding molten material from the barrel of a plastometer under preset conditions of temperature and load. For melt mass-flow rate, timed segments of the extrudate are weighed and the extrudate rate is calculated in g/10 min and recorded. For melt volume-flow rate, the distance that the piston moves in a specified time or the time required for the piston to move a specified distance is measured to generate data in cm3/10 min. Melt volume-flow rate may be converted to melt mass-flow rate, or vice-versa, if the density of the material is known under the conditions of the test.

 

Specification:
Squeeze out the part of material
Discharge port diameter Φ2.095±0.005mm
The length of the discharge port 8.000±0.025mm
Loaded barrel diameter Φ9.550±0.025mm
Loaded barrel length 152±0.1mm
The diameter of the piston rod head 9.475±0.015mm
The length of the piston rod head 6.350±0.100mm
1, Standard Test load (eight)
Ⅲ 2.160 kg = (0.325 + 1.835 weight)
Ⅷ 21.600 kg, = (0.325 +2 No. 0.875 + 1.835 )
Measurement MFR and MVR and Density
Thermostat range Room temperature - 400°C
Temperature accuracy ±0.2°C
Temperature resolution 0.1°C
Timing range 0.1-999 s
Timing accuracy 0.1s
Cut material way Automatic or manual cut material way
Display Color touch screen
Print Can print the test report
Working conditions Ambient temperature of 10°C -40°C; Relative humidity of 30% -80%
Instrument Dimensions 450×400×610(L × W × H)

Send your inquiry directly to us

Privacy Policy China Good Quality Flammability Testing Equipment Supplier. Copyright © 2017-2025 DONGGUAN YUYANG INSTRUMENT CO., LTD . All Rights Reserved.